Veterinary Anesthesia Machine Instrument
Product Features of ventilator part :
1)The high-precision Vaporizer is designed with a stable concentration output,automatic compensation for pressure, temperature and flow rate. And it is available to choose one kind of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
2)Reliable pneumatic transmission system.
3)Economic and proper configuration of technologies, optional as required by the Customer.
Product Specification of ventilator part :
1)Electronically controlled, pneumatic driven
2) Patient spectrum: pediatric to bariatric
3)Ventilation mode: IPPV, SIMV, MANUAL(BAG)
4)Tidal volume: Adjustment range 0-1500ml
5)I:E ratio: 4:1-1:8
6)Frequency: Adjustment range: 1-100bpm
Under IPPV and Manual mode: 4~65bpm Under SIMV: 1-40bpm
7)Monitoring: VT, MV, BPM, Paw, FiO2, PEEP, Respiratory Frequency, I:E ratio, Ppeak
8)Waveform: Pressure- time, Flow- time
9)Power: AC 220V±10%, 50HZ±2%, built-in battery > 2hrs(rechargeable) 12V
10)Inspiratory time (TI) (at SIMV) - Adjustment range: 0.2 s ~ 6.0 s
11) Pressure trigger: - 20 cmH2O ~ + 20 cmH2O
12)Pressure limit: Patient airway pressure range: -2KPa~10KPa Airway pressure monitoring range: 0~10KPa
13)Safety valve pressure limit: 6Kpa
14)Alarm: High(low) pressure alarm, High(Low) MV alarm, AC power lost alarm
Product Features of Anesthesia part :
The ideal portable anesthesia (analgesia) system support an upgrading not only for super clinical field,but also applicable for animal anesthesiology,gynaecology and obstetrics, stomatology and emergency department. It gives a safe,accurate and convenient support for application combination which will assist with fast response to life care.
Product Specification of Anesthesia part :
1)Operating Panel: English (Standard)
2)Drive: Gas driven
3)Application: Applied for animal
4)Available in trolley
5)Main body High-strength engineering plastic rack, light, beautiful and corrosion resistance
6)Scope of application 2 kg above animal
7)Gas source O2:0.28~0.6MPa N2O:0.28~0.6MPa
8)Flow meter O2:0.05~1.0L/min 1.1~10L/min (O2:0.05~10L/min N2O:0.05~10L/min)
9)(O2,N2O linkage and N2O Seopper) (When using nitrous oxide, oxygen concentration≥25%; When the oxygen pressure<0.2Mpa,the flow of nitrous oxide would be cut off )
10)Flow rate of rapid oxygen supply 25~75L/min
11)Low oxygen pressure alarm it will be alarmed when the oxygen pressure < 0.2MPa
12)Vaporizer it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer.
13)Respiratory circuit work mode :all close, semi-close. semi-open APL:≤12.5kPa
14)Respiratory Bellows bellows for adults,bellows for children tidal volume range:0~1500ml
name |
Animal Anesthesia Machine |
Main technical parameters |
Part of Anesthesia Machine: |
Main body: |
high-strength engineering plastic rack,light, beautiful and corrosion resistance |
Scope of application: |
2 kg above animal |
Gas source: |
O2:0.28~0.6MPa N2O:0.28~0.6MPa |
Flow meter : |
O2:0.05~1.0L/min 1.1~10L/min (O2:0.05~10L/min N2O:0.05~10L/min) |
Flow rate of rapid oxygen supply: |
25~75L/min |
Low oxygen pressure alarm: |
high-strength enigneering plastic rack,light,beautiful and corrosion resistance |
Respiratory circuit: |
work mode :all close, semi-close. semi-open APL:≤12.5kPa |
Respiratory Bellows: |
bellows for adults,bellows for children tidal volume range:0~1500ml |
Part of Ventilator |
Display mode: |
high-definition 5.7" LCD screen display |
Part of Ventilator |
Ventilation mode |
IPPV,SIPPV,SIMV,MANUAL |
Ventilation function |
PEEP,SIGH ,IRV |
Ventilation parameters |
Tidal volume: |
50~1500ml |
Rate: |
2~99bpm |
SIMV rate: |
2~20bpm |
I:E: |
2:1~1:8 |
Inspiratory trigger pressure: |
-10~20cmH2O |
PEEP: |
2~20 cmH2O |
Pressure limit: |
10~60 cmH2O |
SIGH: |
1.5 times the inspiratory time/100bpm |
Security Alarm System |
Oxygen Concentration alarm |
upper limit setting range: |
21%~99% |
low limit setting range: |
18%~80% |
Airway pressure alarm |
upper limit setting range: |
5~60 cmH2O |
low limit setting range: |
0~50 cmH2O |
Minute ventilation volume alarm |
upper limit setting range: |
1~20 L/min |
low limit setting range: |
0~19 L/min |
Tidal volume |
upper limit setting range: |
10~1500ml |
low limit setting range: |
0~1400ml |
Power: |
AC 220V 50Hz |
|
|
|
|
You may also like: